Why is this document important for European manufacturers?
Transactions involving medical devices authorised in the US, such as company takeovers, the sale of product lines or the transfer of rights to a specific device, are common in the industry. For Polish and European manufacturers who buy or sell products with 510(k) clearance, the guidelines clarify issues that had previously been a source of uncertainty. The most common question was: following a change of ownership, is it necessary to go through the FDA procedure again?
A new 510(k) submission is required only in the event of substantial changes
The FDA confirms that the new holder does not need to submit a new 510(k) submission if the device is not substantially altered in terms of its design, components, manufacturing method or intended use. The rationale stems from the regulations: a 510(k) notification is submitted prior to the first placing of the device on the market, and a change of ownership alone does not constitute the first placing on the market.
However, if the new holder makes a substantial change to the device, they must, as a general rule, submit a new 510(k) notification. The assessment of whether a change is substantial is carried out on the basis of the existing FDA guidance on changes to devices and software (Deciding When to Submit a 510(k) for a Change to an Existing Device and its equivalent for software).
The applicant’s name does not change in the FDA database. In the 510(k) Premarket Notifications Database, the original applicant continues to be listed as the ‘applicant’, even after the notification has been transferred. Information on who the current holder is can be found in the FDA’s Device Registration and Listing System.
The FDA explicitly states that a device cannot have multiple 510(k) holders at the same time. According to the agency, this model facilitates regulatory enforcement, the tracking of signals and adverse events, and communication with the FDA. At the same time, the FDA emphasises that the guidance is not intended to restrict the transfer of notifications or existing commercial agreements.
Other entities contractually linked to the holder – namely, contract manufacturers, contract sterilizers, repackagers and relabellers – must use the 510(k) notification number assigned by the FDA when listing the device. The initial importer may fulfil the listing obligation by providing the name and address of the manufacturer, provided that they have not themselves developed the specifications or altered the packaging or labelling.
What must a successor manufacturer do?
- Establishment registration – register the establishment within 30 days of commencing activities subject to the registration requirement (if the entity has not previously been registered).
- New device listing – create a new listing in the FDA’s electronic system. The system assigns a new listing number.
- Correct 510(k) number – provide the original 510(k) number if the device has not changed, or the new number if a new notification has been submitted and cleared.
- GUDID update – the new GUDID number and the amended details of the labeler. If the amended details appear on the label, the update must be submitted no later than the date on which the device is first labelled with the new details. If they do not appear on the label, the deadline is 10 working days from the date of the change.
- Labelling changes – the name and place of business of the manufacturer, packer or distributor, and the Unique Device Identifier (UDI) on the label.
- Annual registration update – between 1 October and 31 December, covering devices with newly obtained 510(k) clearances.
The previous holder also has obligations. If they cease to carry out an activity specified in the listing, e.g. manufacturing, they must update the entry. If they cease to distribute the device, they must remove it from the listing.
A letter to the FDA is not sufficient
The FDA draws attention to a common practice: new holders send the agency information regarding the transfer of a notification and request that the databases be updated. The agency has no such obligation. The owner or operator of the establishment is responsible for complying with the registration and listing requirements.
The consequences of negligence can be serious. Failure to register, to list the device, to update the GUDID or to submit the required notification may result in the device being deemed adulterated or misbranded within the meaning of the FD&C Act. The FDA also points out that it uses the registration system, amongst other things, to promptly notify a company of the need for an immediate product recall. Out-of-date information may therefore delay contact in a critical situation.
510(k) notification documentation
The purchaser may request a copy of the entire 510(k) notification dossier from the previous owner or submit a request under the Freedom of Information Act (FOIA). Some of the documents previously made available under FOIA are published by CDRH in the 510(k) database. In practice, it is advisable to stipulate the handover of the full documentation in the sales contract, as without it, it is difficult to assess whether future changes to the device will require a new notification.
Checklist for a transaction involving a 510(k)
- Stipulate in the contract the transfer of the complete 510(k) submission documentation and design documentation.
- Assess whether any changes planned following the acquisition (component supplier, manufacturing site, labelling) are ‘substantial’ within the meaning of the FDA guidelines.
- Plan for registration and a new listing within the 30-day deadline.
- Agree with the previous holder on a date for updating or withdrawing their listing.
- Update GUDID records and labels (UDI, details of the entity responsible for labelling).
- Update the quality management system procedures relating to the reporting of adverse events and recalls, so that responsibility is clearly assigned.
Summary
The final guidance does not introduce new obligations, but organises the existing provisions of 21 CFR 807, 801 and 830 into a practical set of answers. The key takeaways are as follows: the transfer of a 510(k) notification does not in itself require a new notification; a device always has a single holder; and the new owner is solely responsible for registration, listing and updating the GUDID. Companies planning acquisitions in the US market should take these steps into account as early as the due diligence and contract negotiation stages, as part of their regulatory strategy.
Link to the document:
Transfer of a Premarket Notification (510(k)) Clearance – Questions and Answers