Why are separate guidelines needed?
Regulation (EU) 2017/746 (IVDR) requires that, prior to a device being placed on the market, the level of clinical evidence must be sufficient to demonstrate its performance and safety. In the case of a genetic test for a rare disease or a toxicological test for an unusual poisoning, it is difficult to collect a sufficient number of samples within a reasonable timeframe. Added to this is the low sales volume, which means that many such tests are not commercially viable. Consequently, they are often manufactured by healthcare facilities themselves as in-house devices, pursuant to Article 5(5) of the IVDR.
The MDCG notes that, without common guidelines, manufacturers, notified bodies and regulators might interpret the requirements differently. The aim of the document is therefore to ensure the proportionate application of the regulations: the requirements regarding clinical evidence must be met, but without unnecessarily delaying patients’ access to these tests.
When is a device an ‘orphan IVD’?
The guidelines set out two conditions which must be met cumulatively:
- The device is intended for the diagnosis of a disease, condition or situation affecting no more than 12,000 people per year in the EU.
- At least one of the additional conditions is met: a lack of sufficient alternatives, or an expected clinical benefit compared with available solutions or the current state of knowledge.
The threshold of 12,000 people has been adopted from the US ‘Humanitarian Use Device’ (FDA) category and scaled to the EU population of 447 million. It is the incidence rate that counts, not the number of tests carried out. A screening test may therefore cover far more people than are actually affected by the condition and still qualify as an orphan IVD. Where a device relates to a rare disease as defined by the EU (no more than 5 cases per 10,000 people) and is intended to identify no more than 12,000 patients annually, this is sufficient to meet the epidemiological criterion. Data sources may include, amongst others, the registers of the European Reference Networks, EMA data or Orphanet.
Two caveats are particularly important. Firstly, orphan IVD status does not confer market exclusivity: several CE-marked devices may hold this status for the same condition. Secondly, artificially narrowing the intended use to a subgroup of patients solely to meet the threshold will not be accepted. The identified subgroup (orphan subpopulation) must be clinically justified.
Orphan indication. A device intended for use in large populations may also have a separate ‘orphan’ indication. In such cases, the rules set out in the guidelines apply only to that indication. The extrapolation of data from a non-orphan population, described in detail in Annex 3, plays an important role here.
Less data before registration, more obligations afterwards
This is the most important part of the document. A device may be placed on the market with limited data on clinical performance if the manufacturer demonstrates that:
- it has assessed all available data and identified their limitations;
- the existing data demonstrate compliance with the relevant general safety and performance requirements (GSPR) set out in Annex I to the IVDR, an acceptable benefit-risk ratio and an expected clinical benefit;
- it is neither feasible nor proportionate to obtain further data within a reasonable timeframe prior to placing the device on the market;
- it has appropriate PMPF (post-market performance follow-up) and PMS (post-market surveillance) plans in place to address the missing data;
- users are aware of the device’s status, the data limitations and how to report incidents, e.g. in the instructions for use or in the Summary of Safety and Performance (SSP) for classes C and D.
Limitations must relate only to those parameters for which it is genuinely difficult to collect data. The guidelines provide an example: if there is no shortage of negative samples, specificity must be well documented.
Data not covered by the definition of clinical performance data – that is, data on scientific validity and analytical performance – play a significant role. These may be derived, amongst other sources, from the literature and data on similar tests, in silico modelling, tissue and ex vivo models, contrived samples, and even products intended solely for research purposes (RUO). However, each such source must be justified in terms of its relevance to the specific device.
The price for this flexibility is the PMPF plan, which, according to the MDCG, must be detailed and properly structured. The plan should identify any data limitations, how these will be addressed, the type of data to be collected, the anticipated annual number of samples tested, and the milestones for the analyses. Real-world data is also acceptable, including data from other regulatory systems, whilst taking into account physiological differences between populations.
Notified body: certificate with conditions
The notified body should verify the orphan IVD status as early as possible, preferably at the application review stage. The technical documentation is assessed according to the same principles as for other devices. However, the assessment report should clearly justify why limited data and the proposed PMPF measures have been accepted.
The guidelines encourage the issuance of certificates with specific conditions (Annex VII, Section 4.8 and Article 51(3) of the IVDR). These may, for example, impose an obligation to carry out specific PMPF studies within a set timeframe. If the manufacturer fails to meet these conditions, the certificate may be suspended or withdrawn.
Expert panels: 60 or 90 days for an opinion
Consultation with an expert panel is voluntary and separate from the Performance Evaluation Consultation Procedure (PECP) under Article 48(6) of the IVDR. The Paediatrics and Rare Diseases Panel and the IVD Panel may be involved. The choice of panel is decided by the EMA, which provides the secretariat for these panels. The guidelines provide for three routes:
- Early advice for the manufacturer – prior to the clinical performance study. The MDCG considers this route to be the preferred one.
- Advice at the request of a notified body – particularly where the body has doubts regarding the status of the device.
- Late advice at the manufacturer’s request – exceptionally, during the IVDR transition periods, mainly for legacy devices, provided that the manufacturer manages to update the performance evaluation report before it is assessed by the notified body.
The panel will endeavour to issue an opinion on the status itself within 60 days, and an opinion covering clinical data as well within 90 days. If the notified body disagrees with the panel, it must justify its position in the report.
What does this mean for hospital laboratories?
In-house devices are exempt from most IVDR requirements, including certification, but must comply with the relevant GSPR. The general principles set out in Sections 4–9 and Annexes 3 and 4, including data extrapolation, therefore also apply to establishments manufacturing tests for their own use. Part B, which deals with notified bodies and panels, does not apply to them. The guidance should be read in conjunction with MDCG 2023-1 and MDCG 2025-5.
Summary
MDCG 2026-6 does not lower the requirements of the IVDR, but shifts part of the burden of proof from the pre-market phase to the post-market phase. For manufacturers of tests used in rare diseases, this presents an opportunity to obtain CE marking more quickly. However, this requires a well-justified device status and a PMPF plan that will actually be implemented. The guidelines are not legally binding, but are likely to quickly become a point of reference for notified bodies when assessing such devices.